EXPLAINER

What Is CADR in Air Purifiers? HEPA Ratings Explained

Two numbers decide whether an air purifier cleans your room. Almost every explanation of them online is a paraphrase of a paraphrase, and some of it is simply wrong. This page works from the standards themselves: the AHAM test method, the ISO filter classification, and India's own IS 17531. It also covers the part nobody mentions — the three official sizing formulas disagree by a factor of two.

Updated Aug 2026 Standards-sourced 10 min read

01

The short version

HEPA is a pass mark on a filtration test, not a material. CADR is the volume of fully cleaned air the whole machine delivers per unit of time. The filter grade describes one component; CADR describes the appliance.

THE TWO DEFINITIONS THAT MATTER

HEPA: the US EPA defines it as a mechanical filter removing at least 99.97% of particles at 0.3 microns, "with high efficiency for both larger and smaller particles". Under ISO 29463 the equivalent classes are H13 and H14, tested differently — see below.

CADR: the EPA defines it as removal efficiency for a pollutant multiplied by airflow through the device. AHAM adds that CADR is always a measurement of the unit "as a complete system".

For sizing in an Indian city, aim for roughly 11 m³/h of CADR per square metre of floor. During a bad winter, push that closer to 17 m³/h per square metre. Section 06 shows where both figures come from, and why India's own standard asks for less.

If you have already understood the concepts and want machines, prices and filter economics, that is a different page: the air purifier buying guide.

02

HEPA is a test result, and there are two different tests

This is the single most common source of confusion, and most articles get it wrong by quoting both standards as if they were one.

The American definition comes from the Department of Energy and the EPA: 99.97% capture of particles at 0.3 microns. The European definition comes from EN 1822, now internationalised as ISO 29463. It classifies filters by their efficiency at the most penetrating particle size, whatever that size turns out to be.

The two are not interchangeable. H13's 99.95% looks weaker than the American 99.97%, but it is measured at the filter's own worst case, so it is the harder bar to clear. Writing "H13 means 99.97%" mixes the numerator of one standard with the denominator of another.

ISO 29463 and EN 1822 filter classes, efficiency at MPPS, and the marketing labels each is sold under
ISO 29463 EN 1822 Group Efficiency at MPPS Typically sold as
ISO 15 EE11E — not HEPA≥ 95%"HEPA-type", "95% HEPA"
ISO 25 EE12E — not HEPA≥ 99.5%"HEPA-grade", "99% HEPA"
ISO 35 HH13H — HEPA≥ 99.95%"True HEPA", "medical grade"
ISO 45 HH14H — HEPA≥ 99.995%"H14 medical grade"
ISO 55 UU15U — ULPA≥ 99.9995%Cleanrooms, not homes

Scroll the table sideways to see all columns.

ISO 29463 covers efficiencies from 95% up to 99.999995%. Only Groups H and U are HEPA or better. Everything in Group E is, in the standard's own phrasing, an efficient particulate air filter — near-HEPA, and explicitly not the same thing.

03

The 0.3 micron figure is not what most people think

Almost everyone reads "99.97% at 0.3 microns" as a limit: the filter catches things down to 0.3 microns and smaller particles slip through. The opposite is true. Below that size, filters generally get better.

Why a filter has a worst case in the middle

Filter media is a tangle of fibres, not a sieve. Particles are caught three ways, and each works best on a different size.

  1. 1

    Impaction — big particles

    Heavy particles cannot follow the air as it curves around a fibre. Their momentum carries them straight into it. This works well above roughly one micron.

  2. 2

    Interception — mid-size particles

    A particle follows the airstream faithfully but is wide enough to brush a fibre in passing, and sticks. This depends on particle width, so it fades as particles shrink.

  3. 3

    Diffusion — tiny particles

    Very small particles are knocked about by air molecules and wander across streamlines at random. The smaller they are, the more they wander, and the more often they hit a fibre.

Impaction and interception fade as particles shrink. Diffusion only starts to bite once they are very small. In between sits a size that neither mechanism handles well — the most penetrating particle size, or MPPS. That is the filter's worst case, and it is why efficiency rises again on both sides of it.

THE MYTH, STATED PLAINLY

"HEPA can't catch anything smaller than 0.3 microns, so it is useless against viruses." This is wrong, and it shaped bad advice during the pandemic. The EPA's own definition of HEPA ends with the words "with high efficiency for both larger and smaller particles". A 0.1 micron particle is captured more reliably than a 0.3 micron one, not less.

Where the real worst case sits

ISO 29463 states that for micro-glass filter media, the MPPS is usually between 0.12 and 0.25 microns — below the American test size. A 1982 Lawrence Livermore National Laboratory study went further. It calculated the true maximum penetration at 0.21 microns. About seven times more particles get through there than at the 0.3 micron certification size.

That is the practical reason the European approach exists. Testing at 0.3 microns measures a filter slightly to the easy side of its own weak point. Testing at MPPS finds the weak point first, then measures there.

04

Why "HEPA-type" is a different class of product

The efficiency gap between E11 and H13 is easy to quote and easy to shrug off. Ninety-five percent against 99.95% sounds like a rounding difference to most buyers. The gap that actually matters is in how the two are verified.

WHAT ISO 29463 ACTUALLY REQUIRES

Group H (HEPA): filters "are individually tested and their efficiency is determined at MPPS". Each one is then leak tested as a finished item.

Group E (sub-HEPA): efficiency "is determined by statistical sample testing only". The standard then states it outright: "Group E filters cannot and shall not be leak tested."

So a genuine H13 is a filter that was individually measured and individually checked for leaks before it shipped. An E11 sold as "HEPA-type" belongs to a class where a batch is sampled and the seal is never checked at all.

This matters more than the percentage, because air takes the path of least resistance. A perfect H13 sheet in a housing that leaks around the edges performs like a much lower grade, and nothing on the box will tell you. It is also the deeper reason to trust CADR over a filter label: CADR is measured on the assembled machine, leaks included.

H14 IS NOT THE UPGRADE IT LOOKS LIKE

Denser media resists airflow. The same fan pushed through an H14 element moves less air. The finished machine can end up with a lower CADR, more noise and higher power draw than its H13 version. Cleanrooms need H14 because they are chasing absolute particle counts. A bedroom is chasing a clean-up rate, which is CADR. Class first, then CADR — and between two machines, take the higher CADR.

05

What CADR actually measures

CADR is not a manufacturer's claim about airflow. It is the output of a defined experiment, and knowing the experiment tells you what the number can and cannot mean.

The AHAM AC-1 method

The reference method worldwide is ANSI/AHAM AC-1. A sealed chamber of 1,008 cubic feet (28.4 m³, about 10.5 × 12 × 8 ft) is filled with a test aerosol. Concentration decay is measured twice: once with the purifier off, once with it on. The difference between the two decay rates, multiplied by the chamber volume, is the CADR.

CADR in CFM = (decay rate with cleaner on − natural decay) × 1,008

ANSI/AHAM AC-1 test method

Subtracting the natural decay is the important part. Particles settle on their own, and without that subtraction an empty chamber would score a CADR. The method credits the machine only with what it actually removed.

Three pollutants, three numbers

AC-1 runs the test with three standardised aerosols, each standing in for a size band found in real homes.

The three AHAM AC-1 test aerosols and the particle size ranges they represent
Test aerosol Size range Stands in for
Tobacco smoke0.09 – 1.0 µmCooking fumes, candle soot, combustion haze
Fine dust0.5 – 3.0 µmRoad dust, talc-grade household dust
Pollen0.5 – 11.0 µmPollen, larger allergens

Scroll the table sideways to see all columns.

Smoke CADR is the one that matters for pollution. It covers the smallest, hardest particles, and it is the figure every room-size formula is built on. AHAM chose it because those particles are also the ones drawn deepest into the lungs. AC-1-2019 added a separate PM2.5 CADR. It is the geometric mean of the smoke result from 0.1 to 0.5 microns and the dust result from 0.5 to 2.5 microns.

THREE THINGS THE TEST FIXES IN PLACE

It runs at maximum fan speed. AHAM tests the highest setting because "low" and "medium" are not comparable between machines. AHAM also states that running slower cuts both the clean air delivered and the room coverage, at least in proportion to the speed reduction.

It uses a new filter. Nothing in the rating reflects a filter six months into an Indian winter.

It has a ceiling. The maximum CADR that AC-1 can measure reliably is 450 CFM, roughly 765 m³/h. Above that, published figures are extrapolations rather than measurements.

One distinction worth holding on to: CADR is not airflow. Both are quoted in CFM, and AHAM warns against confusing them. Airflow is how much air the fan moves. CADR is how much of that air came out genuinely clean. A machine can move a great deal of air very badly.

06

CADR to room size: three standards, three answers

This is where published guidance quietly falls apart. Three official formulas exist, all defensible, and for the same bedroom they differ by more than a factor of two. Nobody tells you which one their number came from.

The common foundation

All three ask the same question: how much clean air per hour holds indoor pollution at a target level, given that dirty air keeps leaking in? The answer is expressed as equivalent air changes per hour, or eACH — how many times per hour the machine could replace the room's entire air volume with clean air.

The three formulas differ in only two assumptions: how much polluted air leaks in, and how clean you insist the room stays.

The three published room-sizing rules, their assumptions and their outcomes
Rule eACH Pollution held down by Assumes intrusion of
IS 17531:2021 (India)3.280%0.6 air changes/hour
AHAM AC-1 (reference)4.880%1.0 air change/hour
AHAM heavy smoke7.586%1.0 air change/hour, higher outdoor load

Scroll the table sideways to see all columns.

AHAM's 4.8 figure is not arbitrary. It falls out of solving the steady-state balance for an 80% reduction, assuming one air change per hour of unfiltered intrusion. The result is 4.816, rounded to 4.8. That yields the rule printed on the AHAM seal: room area in sq ft = 1.55 × smoke CADR in CFM. The method has been verified by NIST and recognised as reasonable by the US Federal Trade Commission.

India's IS 17531:2021 runs the same steady-state derivation with Indian assumptions: a 2.5 m ceiling, ventilation of 0.6 air changes per hour, and a deposition rate of 0.2 per hour. Its Annex C lands on a much simpler formula — floor area in m² = 0.125 × CADR in m³/h. That implies 3.2 eACH, and it is roughly 48% more generous than AHAM for the same machine, almost entirely because it assumes less dirty air leaking in.

What that means for a real room

Minimum CADR in m³/h, by the three rules. Numbers are computed from the source formulas, converting at 1 CFM = 1.699 m³/h.

Minimum CADR in cubic metres per hour by room size under the Indian standard, the AHAM reference rule and AHAM heavy-smoke guidance
Room IS 17531 (3.2 eACH) AHAM (4.8 eACH) Bad-air days (7.5 eACH)
100 sq ft 9 m² 75 110 170
120 sq ft 11 m² 90 130 205
150 sq ft 14 m² 110 165 255
200 sq ft 19 m² 150 220 340
250 sq ft 23 m² 185 275 425
300 sq ft 28 m² 225 330 510
400 sq ft 37 m² 295 440 680

Scroll the table sideways to see all columns.

Read the 120 sq ft row. A standard Indian bedroom needs 90, 130 or 205 m³/h depending purely on which body's assumptions you accept. All three are published standards. None is wrong.

WHICH COLUMN SHOULD AN INDIAN BUYER USE?

The assumption that decides it is how much polluted air leaks in. India's 0.6 air changes per hour describes a closed room in ordinary air. A Delhi winter is not ordinary air: outdoor PM2.5 runs many times the indoor target, so every leak carries a far heavier load.

That is exactly the situation AHAM wrote its heavy-smoke rule for. Use the middle column as your floor and the right-hand column as your target in a polluted city. The right-hand column is also the simplest rule anyone has published: CADR in m³/h ≈ 1.7 × room area in sq ft.

Why buying far more CADR stops helping

Headroom is good, but the returns fall off a cliff. AHAM publishes the curve, and it is worth seeing before you pay for a much larger machine.

Pollutant reduction achieved at each level of equivalent air changes per hour
eACH delivered 1 2 3 4.8 6 8 10
Pollution removed45%62%71%80%83%87%89%

Scroll the table sideways to see all columns.

Going from 1 to 4.8 eACH nearly doubles the pollution removed. Going from 4.8 all the way to 10 adds nine points. Beyond roughly 7 or 8 eACH you are mostly buying noise. Spend the headroom on running a bigger machine at medium speed instead, which is quieter and gets you the same clean-air rate.

For the working table with ceiling heights applied, and what to do once you have your target number, see the CADR sizing section of the buying guide. Its figures use a round 5 air changes per hour and land within about 3% of AHAM's exact 4.816.

07

Why the coverage area on the box is bigger than any of this

Now the numbers make sense. A coverage area is not measured — it is calculated from the CADR, and whoever does the calculating picks the eACH. Choose a low one and the same machine covers a much larger room on paper.

AHAM addressed this directly in a white paper published in January 2026, and the wording is unusually blunt for an industry body.

It is noted in the marketplace that one eACH is often used by many air cleaner manufacturers. However, we have not found evidence that there are manufacturers stating that the 1 eACH room size reduces only 45% of the pollutants.

AHAM, Consumer Room Air Cleaners and Equivalent Air Changes per Hour, January 2026

AHAM recommends that no coverage claim be based on less than 2.2 eACH. Where a manufacturer departs from the AC-1 method, it says the resulting pollutant reduction "should be clearly communicated for transparency to the consumer".

Running the check yourself

You can reverse any coverage claim in one step. For an 8 ft ceiling:

eACH = smoke CADR in CFM × 60 ÷ (8 × claimed area in sq ft)

AHAM AC-1 Annex E, rearranged

Take a popular India-market unit rated 360 m³/h and advertised at 462 sq ft. That CADR is about 212 CFM, so the implied eACH is roughly 3.4, which corresponds to about 74% pollution removal rather than AHAM's 80%. Under the AHAM rule the same machine suits about 328 sq ft; on a bad-air day, about 212 sq ft.

Notice what that is not. The claim is not invented — it sits close to what India's own IS 17531 formula would give for that machine, around 484 sq ft. It is a real calculation against a softer assumption than a Delhi winter deserves.

ONE CAVEAT ON CROSS-STANDARD MATHS

Most India-market units publish a particulate CADR tested under China's GB/T 18801, not an AHAM smoke CADR. The chambers and test aerosols differ, so converting between them is close but not exact. Treat the comparison above as indicative rather than a lab result.

08

India has its own standard, and it asks for more than most listings show

IS 17531:2021, "Portable Electric Indoor Air Purifier — Specification", was published by the Bureau of Indian Standards in April 2021. It is barely mentioned in Indian air purifier coverage, and it is the most directly useful document a buyer here can know about.

What it requires to be marked on the product

Clause 10 requires the following to appear permanently and legibly on the unit, and again on the packaging box.

  • The CADR for cigarette smoke or salt aerosol, at maximum mode
  • The sound power level in dB(A), at maximum mode
  • The energy grade: Grade I, Grade II or Grade III
  • Power input in watts, rated voltage and frequency

The CADR for every test pollutant must additionally appear in the product manual.

The noise limits nobody quotes

Clause 5.3 ties a noise ceiling to the CADR band, tested to IEC 60704-1 or ISO 3743-2. A louder machine is only permitted if it is also doing more work.

IS 17531:2021 maximum sound power level by CADR band
CADR (m³/h) Max sound power level
Under 15055 dB(A)
150 – 30061 dB(A)
300 – 45066 dB(A)
Above 45070 dB(A)

Scroll the table sideways to see all columns.

Do not compare these to the dB figure in an ad. Sound power level and the sound pressure level brands usually quote are different quantities measured different ways. A "52 dB" marketing claim and a 61 dB(A) sound power limit are not in conflict, and neither can be converted into the other without the test conditions.

The energy grade

Clause 8 grades a purifier on clean air delivered per watt, written as CADR divided by power input in m³/W·h.

  • Grade I — 10 or above
  • Grade II — 6 to 10
  • Grade III — below 6

You can compute it yourself from any spec sheet: divide the CADR in m³/h by the rated wattage. A 360 m³/h machine drawing 33 W scores about 10.9, which is Grade I.

THE CATCH: IT IS VOLUNTARY

Clause 10.3 says products conforming to the standard may be certified and carry the Standard Mark. Nothing compels a brand to test against IS 17531 at all. That is precisely why so many Indian listings publish no CADR, no noise figure and no grade. A listing that does publish all three has told you something real about itself.

IS 17531 also has no ozone clause. If you are weighing an ioniser, plasma or UV feature, use California's benchmark. Electronic air cleaners there must emit no more than 0.050 ppm of ozone to be sold. India has no equivalent rule, so CARB certification is the best available proxy.

09

What CADR cannot tell you

CADR is the best single number available. It is still one number about particles, produced in a sealed chamber, on a new filter, at full fan speed.

NOT MEASURED

Gases, odours and VOCs

The EPA states that current test standards rate CADR only for particle removal. AHAM confirms no universal method for gas or VOC performance has been agreed yet. A CADR figure says nothing about formaldehyde, cooking smells or paint fumes.

NOT MEASURED

Viruses and bacteria

AHAM makes no representation about viral or bacterial load, and states no peer-reviewed test procedure for it exists anywhere. Viruses are 20 to 200 nanometres, so filter media does capture them — but CADR is not the evidence for that claim.

CHANGES OVER TIME

Filter life

Every CADR is measured on a new filter. Nothing in the number anticipates loading. Units tested to China's GB/T 18801 also carry a CCM rating, which does describe lifetime capacity.

DEPENDS ON YOU

Your actual room

The chamber is sealed and its air is deliberately kept well mixed. A real room has open doors, furniture blocking intakes and corners the air never reaches. Placement can undo a large part of the rating.

None of this makes CADR unreliable. It makes it specific. It answers one question precisely: how fast does this machine remove particles from well-mixed air. It is also the only number on the box produced by a method replicated across laboratories for three decades.

The air purifier buying guide picks up from here: filter economics, placement, running costs, and which machines publish honest specs.

10

Frequently asked questions

What is CADR in an air purifier?

CADR stands for Clean Air Delivery Rate. It is the volume of fully cleaned air a purifier delivers, measured in cubic feet per minute or cubic metres per hour. The US EPA defines it as the removal efficiency for a pollutant multiplied by the airflow through the machine, so it captures the fan and the filter in one number. A weak fan with an excellent filter and a strong fan with a mediocre filter can score the same. That is why CADR beats the filter grade as a single comparison figure.

What is a good CADR rating for my room?

There is no single answer, because three published standards use different assumptions. India's own standard, IS 17531:2021, allows roughly 8 m³/h of CADR per square metre of floor. The American AHAM AC-1 method is stricter at about 11 m³/h per square metre. AHAM's heavy-smoke guidance, the closest analogue to a Delhi winter, asks for about 17 m³/h per square metre. For a 120 sq ft bedroom that spread runs from 90 to 205 m³/h. In a polluted Indian city, size to the higher end.

Is a HEPA filter 99.97% or 99.95% efficient?

Both figures are correct, because they come from two different tests. The US Department of Energy and the EPA define HEPA as capturing at least 99.97% of particles at 0.3 microns. The European EN 1822 standard and its international successor ISO 29463 instead test at the most penetrating particle size, and set H13 at 99.95% and H14 at 99.995%. The European figure looks lower but comes from a harder test. Quoting 99.97% as the H13 number mixes the two standards.

What is the difference between True HEPA and HEPA-type filters?

The percentage gap is only half of it. Under ISO 29463, HEPA filters are Group H: every single filter is individually tested for efficiency and then leak tested. Filters below that bar are Group E, and the standard states plainly that Group E filters cannot and shall not be leak tested. So a "HEPA-type" or "E11" filter is not simply a slightly weaker HEPA. It belongs to a class that is sampled in batches rather than individually verified, and whose seal is never checked.

Why is 0.3 microns used to test HEPA filters?

Because it is close to the hardest particle size to catch, though not exactly it. Large particles are caught by impaction and interception; very small particles wander onto fibres by Brownian diffusion. Between those two effects sits an efficiency minimum. ISO 29463 puts that minimum at 0.12 to 0.25 microns for micro-glass media. A 1982 Lawrence Livermore study calculated the true worst case at 0.21 microns, where penetration is about seven times higher than at the 0.3 micron test size.

Do HEPA filters catch particles smaller than 0.3 microns?

Yes, and generally better than they catch 0.3 micron particles. The EPA's own definition describes HEPA as 99.97% efficient at 0.3 microns "with high efficiency for both larger and smaller particles". Below the efficiency minimum, Brownian motion makes tiny particles collide with filter fibres more often, so capture improves as they get smaller. The widespread belief that HEPA is a sieve with 0.3 micron holes is wrong, and it led to real policy errors during the pandemic.

Why does the coverage area on the box look so generous?

Because coverage is a calculation, not a measurement, and the manufacturer chooses the assumption. AHAM sizes rooms at 4.8 equivalent air changes per hour, which holds pollution at 80% below the unfiltered level. In its January 2026 white paper, AHAM notes that many manufacturers instead size at 1 air change per hour. It adds that it has found no manufacturer disclosing that this removes only 45% of pollutants. The number is not usually fabricated. It is computed against a far softer target.

Is CADR measured at the highest fan speed?

Yes. AHAM tests at the highest speed setting as a uniform condition, because "low" and "medium" mean different things on different machines. AHAM also states that running at a lower speed reduces both the clean air delivered and the room coverage, at least in proportion to the speed reduction. Since most people run a purifier on medium overnight for noise reasons, the rated CADR is a ceiling rather than the figure you actually live with.

Does CADR tell me anything about gases, odours or viruses?

No. The EPA states that current test standards rate CADR only for particle removal, and AHAM confirms there is no accepted method yet for rating gas or VOC performance. On viruses, AHAM is explicit that it makes no representation about reducing viral loads and that no peer-reviewed test procedure for this exists anywhere. Viruses are 20 to 200 nanometres, so filter media does capture them, but a CADR figure is not evidence of it.

Does India have its own air purifier standard?

Yes. IS 17531:2021, "Portable Electric Indoor Air Purifier — Specification", sets the Indian test method and marking rules. It requires the smoke CADR, the sound power level in dB(A) and an energy grade of I, II or III to be marked on both the unit and the packaging. Its room-size formula assumes a 2.5 m ceiling and 0.6 air changes per hour of ventilation. Certification against it is voluntary, which is why many listings still publish no CADR at all.

Sources

Every figure on this page comes from a standards document, a regulator or a national laboratory. No secondary summaries are cited.

You may also like

More guides and reviews to help you decide.

BUYING GUIDE
Air Purifier Buying Guide India 2026: CADR and HEPA

How to choose an air purifier in India: demand a real H13 filter, size the CADR to your room with actual math, and budget for filters — not electricity.

Read more →
REVIEW
Best Air Purifier in India 2026

10 best air purifiers in India for 2026 across budget and premium, compared on HEPA grade, CADR, room coverage, noise and filter cost.

Read more →
REVIEW
Best Car Air Purifiers in India 2026

Find the best car air purifier in India 2026. Compare Philips & more with research-based reviews, prices & pros and cons.

Read more →
EXPLAINER
Inverter Technology Explained: What It Actually Saves

Inverter means variable speed, and the BEE gazette defines it as five or more capacity steps. What it saves, checked against real 2026 ISEER labels and prices.

Read more →
EXPLAINER
BEE Star Rating Explained: What It Means in India

What a BEE star rating actually measures, why a 5-star AC and a 5-star fridge are not comparable, which appliances must carry the label, and how to use it.

Read more →
EXPLAINER
Robot Vacuum Navigation - LiDAR vs Camera Mapping, Mopping

Compare LiDAR and camera mapping in robot vacuums. Learn about mopping effectiveness, consumable costs, and which navigation technology works best.

Read more →
Found this helpful? Share it
Naveen Bhavnani
Naveen Bhavnani
Founder & Research Lead

Technologist and product researcher. Every recommendation is backed by aggregating thousands of Amazon reviews, comparing specs across brands and tracking price trends — so you don't have to.

More about how we research →